-
When I close my regulation
knob, why doesn't the vacuum completely shut off?
Most likely, the quad ring has
been damaged. Replace the quad ring and recheck.
-
Do you have fittings to adapt
your regulators to our current system?
We have patient and wall fittings
for all types of connections. They are sold separately but can be
attached to the regulator before shipping unless otherwise specified.
-
What are the specific design
benefits of the pediatric regulators?
The Pediatric Regulators have a
range of 0-100 mm Hg, as opposed to 0-200 mm Hg. They are designed
without the option of applying full line vacuum. There is no
"Line" mode on either of the models that feature selector
valves. As with all Boehringer regulators, their construction allows
many methods of sterilization (see below), which is a key feature for
infection control concerns.
-
How do you disinfect the regulators?
One of the most valuable safety
features of Boehringer Laboratories' Suction Regulators is their
all-metal construction. This construction allows many different types
of disinfection methods: Autoclave, ethylene oxide gas and
Sterrad®. It is important to clean the regulator components prior
to exposure to disinfection cycles.
-
Why don't you set the timing
to the industry standard: 15 and 7?
The timing of Boehringer
Intermitting Regulators is factory-adjusted to approx. 4 seconds off/
8 seconds on at a typical hospital supply. The "ON " time
should be a short as necessary to provide effective drainage. Time
settings in a clinical environment are sometimes adjusted to 8-10
seconds "ON".
Boehringer regulators can be
easily adjusted by the clinical staff should a longer time cycle be
clinically indicated. Competitive models are not easily adjusted by
the clinical staff and are typically set-up to accommodate the
longest "ON" time that is expected.
-
I am turning the high-pressure
regulator up as high as it goes, but it stops at 550 mm Hg. Why won't
it go to 760 mm Hg?
The maximum indicated output of
the high-pressure regulator cannot exceed the vacuum available at the
wall outlet. In this scenario, the system itself is only providing
550 mm Hg vacuum.
-
What do I do if the needle on
the gauge appears to be stuck and won't go back to zero?
Minor adjustments (<40 mm Hg)
are made by removing the plug in the window of the gauge and
adjusting the zero position of the pointer with a screwdriver. Major
adjustments (>40 mmHg) require factory calibration.
-
Why do the 3700 series have a
different looking gauge than any of the others I have seen?
Our patented linear design has
distinct advantages for accuracy and durability. It is a rugged, yet
simple mechanism that never requires calibration, will not travel
beyond zero and will provide consistently accurate readings. The
linear design provides a very robust gauge that and is intended to
tolerate a wide range of abuse in the clinical environment.
-
Who performs repairs on
Boehringer equipment?
Because of our personal
commitment to quality, our factory is the only authorized service
center for our suction regulators and other instruments. Contact us
at 800-642-4945 for a Return Authorization Number.
Ship to:Boehringer Laboratories, Inc.
Attn: Repair Department, 500 East
Washington Street, Norristown, PA 19401
Please be sure to mark the RA# on
the outside of the package.
-
How long will it take if I
send a regulator in to the factory for repair?
Generally, our repair turnaround
for products under warranty is 5 days.
-
Can I buy Boehringer
regulators through a distributor?
You may order direct from the
factory or, if you prefer, you may call the factory for a distributor
in your area.
-
What is the warranty on your
suction regulators?
Our 3700 Series regulators carry
a 10 YEAR Warranty.
Our 7700 Series regulators carry
a 5 YEAR Warranty on the body and 1 year warranty on the gauge.
-
What, if anything, could I be
charged for in the next 10 years?
We do charge a fee for
decontaminating devices that are sent in for repair. You should
always be sure to clean the inside of the regulator before shipping.
Also, evidence of abuse may result in charges for repair.
-
Do you have a fee for doing
repair estimates? Is this fee refundable if we repair the unit?
We charge a $25 fee for our
repair estimates. We will always call with an estimate before
performing any repair that is not covered under warranty. If the
customer approves the repair, the estimate fee is waived.
-
Are you on Premier Contract?
At this time, we are not under
contract with Premier.
-
What does it mean when you say
that your intermitting regulators "vent to atmosphere"?
The design of Models 7702L, 3704,
and 3714 utilizes large ports to effectively return the collection
circuit to atmosphere, thus allowing fluid backflow to irrigate the
catheter tip.
-
The ASTM recommendation for
gastric drainage is 10 LPM, do Boehringer regulators meet this recommendation?
Boehringer intermitting
regulators will exceed a 10 LPM flow rate.
The reason for the ASTM
recommendation is based on more of a historical convention than a
clinical need. The older Wagensteen apparatus that were employed in
gastric drainage were typically low air flow devices. This low air
flow has been adopted by some manufacturers.
Presumably, there is some concern
that rapid decompression of the gastric space creates a comfort issue
for the patient. In the twelve or more years that Boehringer has been
selling Intermitting Regulators, there has never been a complaint of
patient discomfort.
There have been theoretical
concerns raised such as this. In a clinical setting, the fluid flow
through the catheter is always the rate-limiting characteristic. Any
decompression that could be imposed upon the gastric space will be
significantly affected by the much slower flow of viscous fluid in
the collection tube or catheter. We believe that a high flow to the
canister should affect more effective drainage from the space since
the canister will reach the suction setting quicker thus initiating
the movement of fluid through the line sooner.
-
Is annual lubrication of your
regulators recommended?
We recommend that each
institution determine a maintenance schedule based upon an initial
annual inspection and verification of function. This is a very
typical and reasonable recommendation. We also instruct how to clean
and service regulators that become contaminated. A part of this
servicing is lubrication. Obviously, this would only be done in
instances where a significant event occurred such as complete fouling
of the unit with aspirate. Please note that this capability does not
exist on many other competitive products.
-
When I first set up a
regulator in the intermittent mode, I find the high pressure limit
can be exceeded as much as 4 times the setting I have designated, why
is this happening?
What you describe here is most
likely an overshoot of the indicated output when the Intermitter
cycles on. This occurs when there is no collection circuit attached
to the regulator and the inlet to the regulator is occluded.
Attaching a canister to the regulator inlet should prevent this occurrence.
-
We often hear Boehringer refer
to "Droop." What is droop and how is it relative to
promoting Boehringer Suction Regulators?
Accuracy of regulation depends
primarily on gauge accuracy and ability to provide a consistent level
of vacuum under changing flow conditions, which we measure and
specify as "droop".
"Droop" is the
variation in the vacuum as flow changes from a flowing condition to
an occluded condition. We measure droop as the change in vacuum
measured from full flow to dead-ended on a typical collection circuit
with a 14 French catheter.
In other words, with a regulator
set at 100 mm Hg, occlude the 14 Fr. catheter. The change in gauge
indication is defined as droop.
Involuntary pneumatic biopsy, or
tissue damage, can occur when high levels of vacuum are applied to
delicate tissue. With a Boehringer regulator, you can depend on very
low "droop" compared to our competitor's models.
Our regulators are checked on the
assembly line to meet a specification of less than 10 mm Hg droop at
a 100 mm Hg setting. We have made a sizeable investment in regulator
design to achieve this safety feature.
The practical application of this
feature is simply stated: An evaluation of a regulator's droop allows
one to determine whether the device is truly "regulating"
or if it is functioning as a needle valve. A safe and reliable
regulator should regulate to its set position regardless of pressure changes.